Fire Science and Engineering 2017;31(2):113-118.
DOI: https://doi.org/10.7731/KIFSE.2017.31.2.113    Published online April 30, 2017.
산소소생기 Oxylator EM-100, MicroVenT CSI-3000, OXY-LIFE II의 환기량과 기도내압 비교 - 마네킨 연구 -
심규식, 김은미, 노상균
1나사렛대학교 응급구조학과
2나사렛대학교 응급구조학과
3선문대학교 응급구조학과
Comparison of the Ventilatory Volume and Airway Pressures using Oxylator EM-100, MicroVenT CSI-3000, OXY-LIFE II -A manikin study -
Shim Gyu-Sik, Kim Eun-Mee, Roh Sang-Gyun
1Dept. of Emergency Medical Technology, Korea Nazarene Univ.
2Dept. of Emergency Medical Technology, Korea Nazarene Univ.
3Dept. of Emergency Medical Services, Sunmoon Univ.
요약
이 연구는 산소소생기 Oxylator EM-100, MicroVenT CSI-3000, OXY-LIFE II을 이용하여 인공호흡을 시행할 때 환기량과 기도내압을 비교 분석하였다. 실험기간은 2017년 2월 13일로 수집된 자료는 SPSS 18.0을 이용하여 분석하였다. 연구결과 평균 환기량은 Oxylator EM-100이 551.44 ml (${pm}18.70$), MicroVenT CSI-3000은 527.26 ml (${pm}17.89$), OXY-LIFE II는 369.46 ml (${pm}12.30$)의 환기량을 보였고, 평균 기도내압은 Oxylator EM-100이 $11.89cmH_2O$ (${pm}.41$), MicroVenT CSI-3000은 $11.66cmH_2O$ (${pm}.34$), OXY-LIFE II는 $8.02cmH_2O$ (${pm}.25$)로 측정되었다. 이 연구는 현장에서 Oxylator EM-100, MicroVenT CSI-3000, and OXY-LIFE II에 따라 적절한 환기량 전달을 위한 방법을 검증하여 사용방법에 대한 기초자료를 제공하였다.
Abstract
The purpose of this study was to compare the ventilatory volume and airway pressure using Oxylator EM-100, MicroVenT CSI-3000, OXY-LIFE II. The data were obtained from February 13 in 2017 andanalyzed using the SPSS WIN 18.0 program. The results of theventilatory volume showed the Oxylator EM-100 551.44 ml (${pm}18.70$), MicroVenT CSI-3000 527.26 ml (${pm}17.98$), and OXY-LIFE II 369.46 ml (${pm}12.30$). The airway pressure showed the Oxylator EM-100 $11.89cmH_2O$ (${pm}.41$), MicroVenT CSI-3000 $11.66cmH_2O$ (${pm}.34$), and OXY-LIFE II $8.02cmH_2O$ (${pm}.25$). This study will provide the basic data for an appropriate ventilation method by an oxygen supply device including Oxylator EM-100, MicroVenT CSI-3000, and OXY-LIFE II.
Key Words: Airway Pressure, MicroVenT CSI-3000, Oxylator EM-100, OXY-LIFE II, Ventilatory Volume
TOOLS
Share :
Facebook Twitter Linked In Google+ Line it
METRICS Graph View
  • 1 Crossref
  •    
  • 600 View
  • 5 Download
Related articles in Fire Sci. Eng.

Comparison of Ventilatory Volume and Airway Pressures Using Oxylator EM-1002015 October;29(5)



ABOUT
BROWSE ARTICLES
EDITORIAL POLICY
AUTHOR INFORMATION
Editorial Office
Room 906, The Korea Science Technology Center The first building, 22, Teheran-ro 7 Gil, Gangnam-gu, Seoul, Republic of Korea
Tel: +82-2-555-2450/+82-2-555-2452    Fax: +82-2-3453-5855    E-mail: kifse@hanmail.net                

Copyright © 2024 by Korean Institute of Fire Science and Engineering.

Developed in M2PI

Close layer
prev next